FDA Adverse Event Injury Summary report: N

CAPSUREEPI

MDR report key: 3994035 · Received August 8, 2014

Report

Report Number
2182208-2014-02281
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 6, 2014
Report Date
May 16, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
DTB
PMA / PMN Number
P950024
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ANALYSIS REVEALED THAT BEGINNING (B)(6) 2014, SENSING INTEGRITY COUNTS INCREASED TO 328. THERE IS THE POTENTIAL FOR LEAD TO LEAD INTERACTION OR INSULATION BREACH. T-WAVE OVERSENSING CANNOT BE CONFIRMED. THE RIGHT VENTRICULAR VLEAD INTEGRITY WARNING OCCURRED (B)(6) 2014 FOR HIGH RATE NON-SUSTAINED EPISODES AND SHORT INTERVAL COUNTER.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 6721L-50 X2 LEAD, IMPLANTED (B)(6) 2009; 5866-38M ADAPTOR, IMPLANTED (B)(6) 2009; 4965-35 X2 LEAD, IMPLANTED (B)(6) 2009; 4968-35 LEAD, IMPLANTED (B)(6) 2009. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR OVERSENSING OF SHORT INTERVAL COUNTS (SIC) AND NOISE. THE DEVICE CAN TO COIL MORPHOLOGY SHOWED OVER SENSED BEATS CORRESPONDED TO T-WAVE OVERSENSING (TWOS). THE SENSITIVITY WAS DECREASED. THE LEAD REMAINS IN USE WITH CONTINUED FOLLOW-UP. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471795 CAPSUREEPI ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC, INC. 4968-35

Patients

Seq Age Sex Outcome Treatment
1 00086 YR Required Intervention D314TRG ICD