FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3993980 · Received August 8, 2014

Report

Report Number
2649622-2014-09022
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 26, 2014
Report Date
June 26, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD GRADUAL RISE IN BOTH THRESHOLD AND IMPEDANCE. THE LEAD PROGRAMMING WAS ADJUSTED AND THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD THRESHOLD CONTINUE TO RISE, THE IMPEDANCE WAS LOW, AND OVERSENSING OCCURRED. PORTION OF THE LEAD WAS CAPPED AND THE SUPERIOR VENA CAVA (SVC) COIL REMAINED ACTIVE. A NEW RV LEAD WAS IMPLANTED AS WELL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD GRADUAL RISE IN BOTH THRESHOLD AND IMPEDANCE AS WELL AS HIGH IMPEDANCE. THE LEAD PROGRAMMING WAS ADJUSTED AND THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469960 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00068 YR Hospitalization| R