FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3993967
·
Received August 8, 2014
Report
- Report Number
- 2032227-2014-06751
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- April 14, 2014
- Report Date
- July 11, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
NO BUTTON ERROR ALARM NOTED. THE DEVICE HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE DEVICE HAD CRACKED RESERVOIR TUBE.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
Description of Event or Problem · 1
IT WAS REPORTED THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER WAS UNABLE TO CLEAR THE ALARM. THE ALARM WAS EXPLAINED TO THE CUSTOMER. THE DEVICE WAS EXPOSED TO SWEAT. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469116 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-554LWWH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |