FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 3993957
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-09038
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 20, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED; ANALYSIS REVEALED A FLEX FRACTURE OF THE DISTAL CONDUCTOR. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCK THERAPY DUE TO OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD HAD SEVERAL SHORT INTERVAL COUNTS (SIC) WITHIN A FEW HOURS. THE LEAD WAS EXTRACTED AND A NEW LEAD WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 468929 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00061 YR | Hospitalization| L| R | D284TRK ICD, 4196-88 LEAD, 2088TC LEAD |