FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3993957 · Received August 8, 2014

Report

Report Number
2649622-2014-09038
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 20, 2014
Report Date
May 20, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED; ANALYSIS REVEALED A FLEX FRACTURE OF THE DISTAL CONDUCTOR. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCK THERAPY DUE TO OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD HAD SEVERAL SHORT INTERVAL COUNTS (SIC) WITHIN A FEW HOURS. THE LEAD WAS EXTRACTED AND A NEW LEAD WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468929 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00061 YR Hospitalization| L| R D284TRK ICD, 4196-88 LEAD, 2088TC LEAD