ATTAIN PREVAIL DEFLECTABLE CATHETER DELIVERY SYSTEM
Report
- Report Number
- 9612164-2014-01047
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 20, 2014
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- DQY
- PMA / PMN Number
- K061480
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
PRODUCT EVENT SUMMARY: THE CATHETER WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE MECHANICAL OPERATION OF THE CATHETER PEELING/SLITTING/SPLITTING SHOWED A SPIRAL SLIT AND THE MANIPULATOR WIRE WAS BROKEN. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING IMPLANT, THE SLITTING OF THE GUIDE CATHETER WAS HAMPERED IN SUCH A WAY THAT THE WIRE MESH OF THE CATHETER WAS NOT CUT COMPLETELY. THE PHYSICIAN USED ANOTHER SLITTER TO CUT SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469690 | ATTAIN PREVAIL DEFLECTABLE CATHETER DELIVERY SYSTEM | CATHETER, PERCUTANEOUS | DQY | MEDTRONIC IRELAND | 6227DEF04 | 0007008339 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00047 YR |