ATTAIN ABILITY
Report
- Report Number
- 2649622-2014-09070
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 30, 2014
- Report Date
- May 30, 2014
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND THE OUTER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. IT WAS NOTED THAT THE OUTER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).
IT WAS REPORTED THAT WHILE PREPPING THE LV LEAD FOR IMPLANT THE SCRUB TECHNICIAN FLUSHED THE LEAD, WHICH CAUSED A BALLOONING OF THE SILICONE COATING. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469403 | ATTAIN ABILITY | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 419688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00086 YR |