FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 3993874 · Received August 8, 2014

Report

Report Number
2649622-2014-09070
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 30, 2014
Report Date
May 30, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND THE OUTER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. IT WAS NOTED THAT THE OUTER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE PREPPING THE LV LEAD FOR IMPLANT THE SCRUB TECHNICIAN FLUSHED THE LEAD, WHICH CAUSED A BALLOONING OF THE SILICONE COATING. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469403 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419688

Patients

Seq Age Sex Outcome Treatment
1 00086 YR