SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2014-09084
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 28, 2014
- Report Date
- May 28, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE. IT WAS ALSO NOTED THAT ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE TREND ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS RISING. (B)(4)
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076-52 LEAD, IMPLANTED (B)(6) 2003. (B)(4).
IT WAS LATER REPORTED THAT AN RV LEAD FRACTURE WAS CONFIRMED AND THE LEAD WAS EXTRACTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH THRESHOLDS AND HIGH IMPEDANCE. THE LEAD REMAINS IN USE AND THE PHYSICIAN WANTS THE PATIENT TO RETURN TO THE OFFICE IN ABOUT TWO MONTHS FOR EVALUATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469395 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694758 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00062 YR | Hospitalization| R | D224DRG ICD |