MAXIMO VR
Report
- Report Number
- 3004209178-2014-14570
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 4, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS CONFIRMED THAT A POR OCCURRED ON (B)(6) 2014.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT EVALUATION OF RECORDED DEVICE DATA REVEALED THAT A POWER ON RESET (POR) HAD OCCURRED WITH THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE ICD PARAMETERS WERE CHECKED AND HAD REMAINED AT THE PROGRAMMED SETTINGS. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469398 | MAXIMO VR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | 7232CX |