FDA Adverse Event Malfunction Summary report: N

MAXIMO VR

MDR report key: 3993856 · Received August 8, 2014

Report

Report Number
3004209178-2014-14570
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 4, 2014
Report Date
June 4, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS CONFIRMED THAT A POR OCCURRED ON (B)(6) 2014.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT EVALUATION OF RECORDED DEVICE DATA REVEALED THAT A POWER ON RESET (POR) HAD OCCURRED WITH THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE ICD PARAMETERS WERE CHECKED AND HAD REMAINED AT THE PROGRAMMED SETTINGS. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469398 MAXIMO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO 7232CX

Patients

Seq Age Sex Outcome Treatment
1 00061 YR 694765 LEAD