FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3993852 · Received August 8, 2014

Report

Report Number
2649622-2014-09080
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 26, 2014
Report Date
May 27, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: D284DRG ICD, IMPLANTED: (B)(6) 2012. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPIES DUE TO RIGHT VENTRICULAR (RV) LEAD NOISE OVERSENSING, HIGH THRESHOLDS, HIGH IMPEDANCE AND A POSSIBLE FRACTURE. THE RV LEAD REPROGRAMMED AND THEN EVENTUALLY CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470956 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00057 YR Hospitalization| L| R 5076-52 LEAD