FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 3993848 · Received August 8, 2014

Report

Report Number
2649622-2014-09076
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: ADDR01 IPG IMPLANTED (B)(6) 2008; 310C27 MITRAL VALVE, IMPLANTED (B)(6) 2006. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS FUNCTIONING NORMALLY PRIOR TO A SYSTEM UPGRADE PROCEDURE. HOWEVER, HIGH IMPEDANCE DUE TO POSSIBLE FRACTURE WAS OBSERVED DURING THE PROCEDURE. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469808 CAPSUREFIX DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 4068-45

Patients

Seq Age Sex Outcome Treatment
1 00073 YR Hospitalization| R 5092-52 LEAD