PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2183613-2014-01067
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- April 24, 2014
- Report Date
- June 11, 2014
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Removal / Correction Number
- Z-1661-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; UPPER AND LOWER CASES ARE BROKEN. ANALYSIS ALSO FOUND ONE BATTERY LATCH AND ONE BATTERY RELEASE SPRING ARE DAMAGED. THE HIGH RATE COVER, TWO CONTROL KNOBS, ONE CASE SCREW, ONE SIDE BAIL , AND THE RING ARE MISSING. TWO CONTROL KNOBS, TWO BAIL COVERS, AND RING COVER ARE BROKEN. (B)(4).
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD A DAMAGED CASE. THE EPG WAS RETURNED TO THE MANUFACTURER FOR REPAIR AND CALIBRATION, ANALYZED, AND TESTED OUT OF SPECIFICATION. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469804 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MILACA, INC. | 5348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |