FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 3993754 · Received August 8, 2014

Report

Report Number
2649622-2014-09138
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE INNER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO MIO (METAL ION OXIDATION) WHILE IN VIVO. VISUAL ANALYSIS FOUND INNER INSULATION BREACHED IN-VIVO/ METAL ION OXIDATION. THE BREACHED INSULATION DID CONTRIBUTE TO THE ELECTRICAL COMPLAINTS.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ICD, IMPLANTED: (B)(6) 2003; 419378 LEAD, IMPLANTED: (B)(6) 2003. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND LOW IMPEDANCES. IT WAS FURTHER REPORTED THAT THE RA LEAD EXHIBITED UNDERSENSING/FAILURE TO SENSE, WITH P WAVES OF 0.1 MV. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PATIENT IS PART OF THE SLS CLINICAL STUDY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467385 CAPSUREFIX ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4568-45

Patients

Seq Age Sex Outcome Treatment
1 00065 YR Hospitalization| R 694758 LEAD