CAPSUREFIX
Report
- Report Number
- 2649622-2014-09138
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 18, 2014
- Report Date
- June 18, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE INNER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO MIO (METAL ION OXIDATION) WHILE IN VIVO. VISUAL ANALYSIS FOUND INNER INSULATION BREACHED IN-VIVO/ METAL ION OXIDATION. THE BREACHED INSULATION DID CONTRIBUTE TO THE ELECTRICAL COMPLAINTS.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ICD, IMPLANTED: (B)(6) 2003; 419378 LEAD, IMPLANTED: (B)(6) 2003. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND LOW IMPEDANCES. IT WAS FURTHER REPORTED THAT THE RA LEAD EXHIBITED UNDERSENSING/FAILURE TO SENSE, WITH P WAVES OF 0.1 MV. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PATIENT IS PART OF THE SLS CLINICAL STUDY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467385 | CAPSUREFIX | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 4568-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00065 YR | Hospitalization| R | 694758 LEAD |