FDA Adverse Event Malfunction Summary report: N

ATTAIN OTW

MDR report key: 3993693 · Received August 8, 2014

Report

Report Number
2649622-2014-09183
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 19, 2014
Report Date
May 19, 2014
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 694758 LEAD, IMPLANTED: (B)(6) 2004; 407645 LEAD, IMPLANTED: (B)(6) 2004. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY (B)(4): THE (B)(4) LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE LV (LEFT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE AND ANALYSIS OF THE DEVICE MEMORY INDICATED A LEAD IMPEDANCE OUT OF RANGE ALERT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) AND SUPERIOR VENA CAVA (SVC) IMPEDANCES HAVE RISEN. THE LEFT VENTRICULAR (LV) LEAD IMPEDANCE ALSO ROSE AT THE SAME TIME AS THE RV AND SVC DID. THE RV LEAD HAS A POSSIBLE FRACTURE AND IT WAS CAPPED AND A NEW SINGLECOIL LEAD WAS IMPLANTED. THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472139 ATTAIN OTW PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 419378

Patients

Seq Age Sex Outcome Treatment
1 00071 YR D224TRK ICD