ATTAIN OTW
Report
- Report Number
- 2649622-2014-09183
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 19, 2014
- Report Date
- May 19, 2014
- Manufacturer
- MPRI
- Product Code
- LWP
- PMA / PMN Number
- P010015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 694758 LEAD, IMPLANTED: (B)(6) 2004; 407645 LEAD, IMPLANTED: (B)(6) 2004. (B)(4).
PRODUCT EVENT SUMMARY (B)(4): THE (B)(4) LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE LV (LEFT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE AND ANALYSIS OF THE DEVICE MEMORY INDICATED A LEAD IMPEDANCE OUT OF RANGE ALERT.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) AND SUPERIOR VENA CAVA (SVC) IMPEDANCES HAVE RISEN. THE LEFT VENTRICULAR (LV) LEAD IMPEDANCE ALSO ROSE AT THE SAME TIME AS THE RV AND SVC DID. THE RV LEAD HAS A POSSIBLE FRACTURE AND IT WAS CAPPED AND A NEW SINGLECOIL LEAD WAS IMPLANTED. THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472139 | ATTAIN OTW | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MPRI | 419378 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00071 YR | D224TRK ICD |