FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 3993692
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-09182
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 30, 2014
- Report Date
- June 30, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT UNDER FLUOROSCOPY, THE RIGHT ATRIAL (RA) LEAD WAS FOUND TO HAVE DISLODGED AND WAS NO LONGER SITTING IN THE ATRIUM. DUE TO THE DISLODGEMENT, THE RA LEAD WAS UNABLE TO CAPTURE AT HIGH PACING AMPLITUDE. THE LEAD WAS CAPPED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467287 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00059 YR | Hospitalization| R | D224TRK ICD, 6947-65 LEAD, 4194-88 LEAD |