CAPSUREFIX NOVUS
Report
- Report Number
- 3008973940-2014-00303
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 23, 2014
- Manufacturer
- MEDTRONIC SINGAPORE OPERATIONS
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/ STRETCHING/OVERSTRESS. THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE DURING ATTEMPT TO IMPLANT THE RIGHT VENTRICULAR (RV) LEAD THERE WERE PROBLEMS CROSSING THE RIGHT VENTRICULAR AND IT WAS NOTED THAT THE LEAD WAS HOOKED. THE RV LEAD WAS WITHDRAWN AND EXCHANGED FOR A DIFFERENT LEAD. UPON WITHDRAWAL OF THE RV LEAD IT WAS NOTED THAT THE TIP WAS PARTIALLY EXTRACTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.