FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3993687 · Received August 8, 2014

Report

Report Number
3008973940-2014-00303
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 11, 2014
Report Date
June 23, 2014
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/ STRETCHING/OVERSTRESS. THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DURING ATTEMPT TO IMPLANT THE RIGHT VENTRICULAR (RV) LEAD THERE WERE PROBLEMS CROSSING THE RIGHT VENTRICULAR AND IT WAS NOTED THAT THE LEAD WAS HOOKED. THE RV LEAD WAS WITHDRAWN AND EXCHANGED FOR A DIFFERENT LEAD. UPON WITHDRAWAL OF THE RV LEAD IT WAS NOTED THAT THE TIP WAS PARTIALLY EXTRACTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472134 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC SINGAPORE OPERATIONS 5076-58

Patients

Seq Age Sex Outcome Treatment
1