FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3993684 · Received August 8, 2014

Report

Report Number
2649622-2014-09202
Event Type
Injury
Date Received
August 8, 2014
Report Date
June 17, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. A PROXIMAL PORTION WAS RECEIVED MEASURING 15.5 CM. A DISTAL PORTION WAS RECEIVED MEASURING 49 CM.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD SEVERAL NON-PHYSIOLOGIC NON-SUSTAINED VENTRICULAR TACHYCARDIAS (NST) AND A VERY HIGH NUMBER OF SHORT INTERVAL COUNTS (SIC). THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472137 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00069 YR Hospitalization| R