CAPSUREFIX MRI SURESCAN
Report
- Report Number
- 2649622-2014-09213
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- June 5, 2014
- Report Date
- June 5, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE ANALYST NOTED THAT THE ELECTRICAL RESISTANCE AND CROSS CONTINUITY TEST RESULTS WERE WITHIN SPECIFICATION.
IT WAS REPORTED THAT DURING THE PACING LEAD IMPLANT ATTEMPT THE LEAD WAS UNABLE TO SENSE WHAT LOOKED LIKE SINUS RHYTHM FOR THE PATIENT. AFTER SEVERAL ATTEMPTS BY THE PHYSICIAN TO REPOSITION THE LEAD, IT WAS SUSPECTED THE LEAD HAD BEEN DAMAGED OR KINKED DURING INSERTION. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472665 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR |