FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3993571 · Received August 8, 2014

Report

Report Number
2183613-2014-01073
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 30, 2014
Report Date
May 30, 2014
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE BIOMEDICAL ENGINEER (BE) THAT WHEN THE EXTERNAL PULSE GENERATOR (EPG) WAS POWERED UP, THE DISPLAY CONTAINED NON-ENGLISH WORDS. OF NOTE, THE BATTERY VOLTAGE WAS LESS THAN 6.5 VOLTS. THE BATTERY WAS REPLACED AND THE EPG WAS POWERED UP AS DESIGNED AND WAS PLACED BACK IN SERVICE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467193 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1