FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3993537 · Received August 8, 2014

Report

Report Number
2649622-2014-09244
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 27, 2014
Report Date
May 27, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS 5086MRI LEAD IMPLANTED: 2014-(B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT WITHIN A MONTH OF IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A LOSS OF CAPTURE AT MAXIMUM OUTPUTS. THE LEAD WILL BE REVISED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472239 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R A2DR01 IPG