FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3993481 · Received August 8, 2014

Report

Report Number
2649622-2014-09277
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 10, 2014
Report Date
June 10, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: 8042B BIV IPG IMPLANTED: 2010 (B)(6); 5076-58 LEAD IMPLANTED: 2010 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD DIAPHRAGMATIC STIMULATION IN ALL LV VECTOR CONFIGURATIONS. THE LV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
466710 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419688

Patients

Seq Age Sex Outcome Treatment
1 00075 YR Hospitalization| R 5076-52 LEAD