CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2014-09348
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- February 17, 2014
- Report Date
- June 26, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE OUTER INSULATION OF THE LEAD HAD DEVELOPED A BREACH DUE TO A COSMETIC DEPRESSION THAT HAD DEVELOPED WHILE IN VIVO. VISUAL ANALYSIS OF THE LEAD OBSERVED THE OUTER INSULATION BREACH WITH BLOOD INGRESSION. CONCOMITANT PRODUCT: 407645 LEAD, IMPLANTED (B)(6) 2011. (B)(4).
THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED AND REPLACED DUE TO A SYSTEM UPGRADE. THE LEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469057 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 407652 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00061 YR | Hospitalization| R | ADDRL1 DEVICE |