FDA Adverse Event Injury Summary report: N

PROTECTA XT CRT-D

MDR report key: 3993354 · Received August 8, 2014

Report

Report Number
3004209178-2014-14640
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 9, 2014
Report Date
June 9, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: (B)(4) DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: D314TRG DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT PRODUCT: 419588 LEAD, IMPLANTED: (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS LATER REPORTED BY THE PATIENT THAT THE ¿BATTERY WAS IN PIECES.¿ THE PATIENT ALSO REPORTED THAT A FEW YEARS AGO, THE DEVICE WAS ¿UNHOOKED¿ AND WHEN IT WAS ¿HOOKED UP¿ IT FELT LIKE A ¿BABY KICKING IN THERE.¿ ADDITIONAL INFORMATION HAS BEEN REQUESTED AND NOT YET RECEIVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE TRIGGERED ELECTIVE REPLACEMENT INDICATOR (ERI) EARLY AND THE DEVICE ONLY LASTED 2 YEARS DUE TO HIGH OUTPUTS. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD HIGH IMPEDANCE DUE TO A LEAD FRACTURE. THE DEVICE WAS EXPLANTED AND REPLACED. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468853 PROTECTA XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D314TRG

Patients

Seq Age Sex Outcome Treatment
1 00074 YR Hospitalization| R 6949 LEAD