FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3993330 · Received August 8, 2014

Report

Report Number
2649622-2014-09361
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 26, 2014
Report Date
June 26, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ADDR01, IPG, IMPLANTED: (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION. ELECTRICAL CONTINUITY TEST RESULTS WERE PASSED. THE ANALYST NOTED THAT VISUAL ANALYSIS FOUND OUTER INSULATION BREACHED IN-VIVO/ DEPRESSION AND BLOOD INGRESSION IN LUMEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXPERIENCED HIGH THRESHOLDS AND LOW R-WAVES. THE PHYSICIAN FEELS THIS MAY BE DUE TO SUBCLAVIAN CRUSH. THE RV LEAD WAS EXPLANTED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469758 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00058 YR Hospitalization| R 555445 LEAD