CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2014-09361
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 26, 2014
- Report Date
- June 26, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ADDR01, IPG, IMPLANTED: (B)(6) 2010. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION. ELECTRICAL CONTINUITY TEST RESULTS WERE PASSED. THE ANALYST NOTED THAT VISUAL ANALYSIS FOUND OUTER INSULATION BREACHED IN-VIVO/ DEPRESSION AND BLOOD INGRESSION IN LUMEN.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXPERIENCED HIGH THRESHOLDS AND LOW R-WAVES. THE PHYSICIAN FEELS THIS MAY BE DUE TO SUBCLAVIAN CRUSH. THE RV LEAD WAS EXPLANTED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469758 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00058 YR | Hospitalization| R | 555445 LEAD |