FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3993309 · Received August 8, 2014

Report

Report Number
2649622-2014-09401
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 5, 2014
Report Date
June 5, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: A 5076 LEAD, IMPLANTED: (B)(6) 2007. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCES. THE LEAD WAS CAPPED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, IT WAS NOTED THAT THERE WAS AN INSULATION BREACH AT THE BASE OF THE SUPERIOR VENA CAVA (SVC) COIL CONNECTOR PIN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469577 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00084 YR Hospitalization| R DDBB1D1 ICD