FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3993306 · Received August 8, 2014

Report

Report Number
2182208-2014-02324
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 13, 2014
Report Date
June 18, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND IT TOOK 1 MINUTE AND 45 SECONDS TO BOOT. ANALYSIS DID NOT REPRODUCE OR CONFIRM THE REPORTED FATAL ERROR. ANALYSIS ALSO FOUND THE VVI BUTTON IS INOPERABLE; VVI PRINTED CIRCUIT BOARD ASSEMBLY REPLACED TO RESOLVE. THE UPPER HINGE PLATE IS CRACKED. LEFT ANTENNA FINAL FUNCTIONAL TEST FAILED; LEFT ANTENNA SECURED TO RESOLVE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED FATAL ERROR MESSAGES WERE RECEIVED AND UPON STARTUP THE PROGRAMMER HAS REALLY LONG BOOT. THE PROGRAMMER WAS RETURNED TO THE MANUFACTURER FOR REPAIR, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469580 CARELINK PROGRAMMER, PACEMAKER KRG MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1