SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2014-09415
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 11, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Removal / Correction Number
- Z-0474-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 4076 LEAD, IMPLANTED: (B)(6) 2005, 4194 LEAD IMPLANTED: (B)(6) 2005. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS AND WAS ANALYZED. THE OUTER INSULATION OF THE LEAD WAS BREACHED DUE TO BI/MULTI-LUMEN TUBING VOIDS WHILE IN VIVO. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCES AND AN APPARENT FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470198 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 693565 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00066 YR | Hospitalization| R | D224TRK ICD |