FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3993251 · Received August 8, 2014

Report

Report Number
2649622-2014-09415
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 11, 2014
Report Date
June 11, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0474-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 4076 LEAD, IMPLANTED: (B)(6) 2005, 4194 LEAD IMPLANTED: (B)(6) 2005. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS AND WAS ANALYZED. THE OUTER INSULATION OF THE LEAD WAS BREACHED DUE TO BI/MULTI-LUMEN TUBING VOIDS WHILE IN VIVO. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCES AND AN APPARENT FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470198 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693565

Patients

Seq Age Sex Outcome Treatment
1 00066 YR Hospitalization| R D224TRK ICD