FDA Adverse Event Injury Summary report: N

MAXIMO VR

MDR report key: 3993205 · Received August 8, 2014

Report

Report Number
3004209178-2014-14675
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 28, 2014
Report Date
May 28, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD TRIGGERED ELECTIVE REPLACEMENT INDICATOR (ERI). APPROXIMATELY ONE MONTH LATER, THE PATIENT WAS BROUGHT IN TO CHANGE OUT THE DEVICE, BUT THE BATTERY VOLTAGE WAS NO LONGER AT ERI. THE PROCEDURE WAS CANCELLED. THE DEVICE WAS REPLACED APPROXIMATELY TWO MONTHS LATER AFTER REACHING ERI AGAIN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468905 MAXIMO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO 7232CX

Patients

Seq Age Sex Outcome Treatment
1 00033 YR Hospitalization| R 6947-65 LEAD