SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2014-09477
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- June 6, 2014
- Report Date
- June 6, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
EVALUATION SUMMARY: ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND, HOWEVER THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD, THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT; FULL LEAD RETURNED AND ANALYZED.
IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, AFTER MULTIPLE ATTEMPTS TO REPOSITION THE LEAD, THE HELIX STOPPED DEPLOYING. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472475 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR |