FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3993119 · Received August 8, 2014

Report

Report Number
2182208-2014-02345
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 30, 2014
Report Date
May 30, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; RF (RADIO FREQUENCY) HEAD CABLE HAS EXPOSED WIRES DUE TO DAMAGED INSULATION. ANALYSIS ALSO FOUND THE RF HEAD LENS IS CRACKED AND THE RF HEAD LABEL IS DELAMINATED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RADIO FREQUENCY (RF HEAD) HAS EXPOSED WIRES. THE RF HEAD HAS BEEN RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.