SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2014-09490
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 13, 2014
- Report Date
- June 18, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Removal / Correction Number
- Z-0474-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. (B)(4).
IT WAS REPORTED THAT THE PATIENT RECEIVED SEVEN INAPPROPRIATE SHOCKS DUE TO T-WAVE OVERSENSING (TWOS). THE PHYSICIAN REPROGRAMMED THE RIGHT VENTRICULAR (RV) LEAD SENSITIVITY AND MODIFIED THE DEVICE DETECTION ZONE. THE DEVICE REMAINS IN USE AND A CHANGE OUT WAS SCHEDULED FOR A NEW DEVICE WITH THE TWOS ALGORITHM. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472026 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 693565 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R | D144VRC ICD |