FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3993052 · Received August 8, 2014

Report

Report Number
2649622-2014-09511
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. THE ANALYST COMMENTED THAT HELIX EXTENSION/RETRACTION AND LENGTH TESTING WERE ALL PERFORMED AND ALL RESULTS, INCLUDING ELECTRICAL TESTING, WERE WITHIN SPECIFIED PARAMETERS. BLOOD WAS OBSERVED IN THE SLEEVEHEAD AND ON THE HELIX WHICH MAY CONTRIBUTE TO THE APPEARANCE OF THE HELIX BEING ¿SLOW TO EXTEND¿.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS ATTEMPTED TO IMPLANT BUT NOT USED. THE HELIX EXTENSION MECHANISM WAS SLOW TO EXTEND. THE LEAD WAS ALSO OVERSENSING AFTER PACED EVENTS. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472789 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M55

Patients

Seq Age Sex Outcome Treatment
1 00068 YR