FDA Adverse Event Injury Summary report: N

MYOCARDIAL LEAD

MDR report key: 3993046 · Received August 8, 2014

Report

Report Number
2649622-2014-09534
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
K031274
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: 5071-35 LEAD IMPLANTED: 2013 (B)(6); 4968-35 LEAD 2013 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TWO RIGHT VENTRICULAR (RV) EPICARDIAL LEADS THAT WERE ADAPTED TOGETHER AND WERE OVERSENSING T WAVES. THE LEADS WERE REPROGRAMMED AND REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472787 MYOCARDIAL LEAD ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5071-35

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Required Intervention ADDRL1 IPG