FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3993033 · Received August 8, 2014

Report

Report Number
2649622-2014-09532
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 13, 2014
Report Date
June 13, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 5076-52, LEAD, IMPLANTED: (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. NO ANOMALIES WERE FOUND. THE EXPOSED SVC (SUPERIOR VENA CAVA) DEFIBRILLATION COIL BECAME EXTRINSICALLY FRACTURED DUE TO A CUT. THE INNER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION. VISUAL ANALYSIS INDICATES APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD IMPEDANCE WAS TRENDING LOW. THE LEAD WAS EXTRACTED AND A NEW LEAD WAS IMPLANTED. DURING THE RV LEAD EXTRACTION PROCEDURE, THE LEAD WAS ENTANGLED WITH THE RIGHT ATRIAL (RA) LEAD AND CAUSED THE RA LEAD TO DISLODGE. THE PHYSICIAN WAS UNABLE TO REPOSITION THE LEAD SO THE LEAD WAS EXTRACTED AND A NEW RA LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468474 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00074 YR Hospitalization| R 7272 ICD