FDA Adverse Event Injury Summary report: N

ENRHYTHM MRI

MDR report key: 3992969 · Received August 8, 2014

Report

Report Number
9614453-2014-01931
Event Type
Injury
Date Received
August 8, 2014
Date of Event
April 23, 2014
Report Date
June 17, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, AND RETURNED PRODUCT TESTING FOUND THE DEVICE DID PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY : THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) ONLY LASTED THREE AND A HALF YEARS AND EARLY BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
466956 ENRHYTHM MRI PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND EMDR01

Patients

Seq Age Sex Outcome Treatment
1 00026 YR Hospitalization| R 5086MRI-52 LEA, 5086MRI-58 LEAD