FDA Adverse Event Injury Summary report: N

PROTECTA XT CRT-D

MDR report key: 3992940 · Received August 8, 2014

Report

Report Number
9614453-2014-01936
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 26, 2014
Report Date
May 26, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 86% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT: 5076 LEAD IMPLANTED: 2006-(B)(6); 6931 LEAD IMPLANTED: 2006-(B)(6). (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED DUE TO PREMATURE "FAST" ELECTIVE REPLACEMENT INDICATOR (ERI). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467254 PROTECTA XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND D354TRG

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R 4194 LEAD