SIGMA 300 VDD
Report
- Report Number
- 3004209178-2014-14722
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- April 3, 2014
- Report Date
- May 19, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Removal / Correction Number
- Z-1517-2009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE RETURNED DEVICE INDICATED NORMAL BATTERY DEPLETION. THE RETURNED DEVICE REVEALED LIFTED HYBRID BOND WIRES. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) ENCOUNTERED A CHARGE TIME OUT AND WAS EXPLANTED AND REPLACED. THE ATRIAL LEAD EXHIBITED OVERSENSING, WAS CAPPED AND NOT REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472112 | SIGMA 300 VDD | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | SVDD303 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R | 5038 LEAD |