FDA Adverse Event Malfunction Summary report: N

SIGMA 300 VDD

MDR report key: 3992929 · Received August 8, 2014

Report

Report Number
3004209178-2014-14722
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
April 3, 2014
Report Date
May 19, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1517-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE RETURNED DEVICE INDICATED NORMAL BATTERY DEPLETION. THE RETURNED DEVICE REVEALED LIFTED HYBRID BOND WIRES. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) ENCOUNTERED A CHARGE TIME OUT AND WAS EXPLANTED AND REPLACED. THE ATRIAL LEAD EXHIBITED OVERSENSING, WAS CAPPED AND NOT REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472112 SIGMA 300 VDD PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO SVDD303

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R 5038 LEAD