FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 3992887
·
Received August 8, 2014
Report
- Report Number
- 3008973940-2014-00321
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 28, 2014
- Report Date
- June 6, 2014
- Manufacturer
- MEDTRONIC SINGAPORE OPERATIONS
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING /STRETCHING/OVERSTRESS. THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE LEAD¿S HELIX WAS BLOCKED. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471850 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MEDTRONIC SINGAPORE OPERATIONS | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |