CAPSUREFIX MRI SURESCAN
Report
- Report Number
- 2649622-2014-09645
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THE ANALYST COMMENTED THAT ELECTRICAL RESISTANCE AND CROSS CONTINUITY TEST RESULTS WERE WITHIN SPECIFICATION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE LEAD WAS ATTEMPTED TO BE IMPLANTED IN THE RIGHT ATRIAL (RA) BUT WAS NOT USED. THE LEAD COULD NOT CAPTURE OR SENSE FROM MULTIPLE LOCATIONS. THE PATIENT EITHER HAD VERY FINE ATRIAL FIBRILLATION OR ATRIAL STANDSTILL. THE PHYSICIAN ELECTED TO NOT IMPLANT A LEAD IN THE ATRIAL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471241 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR |