FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 3992824 · Received August 8, 2014

Report

Report Number
3004209178-2014-14761
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS PRODUCT WAS INCLUDED IN A FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT BE DETERMINED IF THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. CONCOMITANT MEDICAL PRODUCTS: 4457 BOSTON SCIENTIFIC COMPETITOR LEAD, IMPLANTED (B)(6) 2009. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INTERROGATION OBSERVED A LEAD WARNING ON THE RIGHT ATRIAL (RA) LEAD FOR HIGH IMPEDANCE TRENDS. IT WAS SUSPECTED THAT THE RA LEAD MAY NOT HAVE BEEN FULLY RE-SEATED INTO THE HEADER DURING A PREVIOUS REPLACEMENT PROCEDURE FOR THE VENTRICLE LEAD. THE IMPLANTABLE PULSE GENERATOR (IPG) DEVICE AND RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS ADDITIONALLY REPORTED THAT THE IPG DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471752 ENRHYTHM DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO P1501DR

Patients

Seq Age Sex Outcome Treatment
1 00084 YR Hospitalization| R