FDA Adverse Event
Injury
Summary report: N
SECURA DR
MDR report key: 3992819
·
Received August 8, 2014
Report
- Report Number
- 3004209178-2014-14753
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 13, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 4592-53, LEAD, IMPLANTED: (B)(6) 2010. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY FOR RAPID VENTRICULAR RESPONSE WITH ATRIAL FLUTTER. THE RIGHT ATRIAL (RA) LEAD IMPEDANCE WAS WITHIN SPECIFICATION PRIOR TO THE SHOCK, BUT IS NOW LOW AND OUT OF RANGE. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470996 | SECURA DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D224DRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00044 YR | Life Threatening | 6944-65 LEAD |