FDA Adverse Event Injury Summary report: N

CAPSURE VDD 2

MDR report key: 3992816 · Received August 8, 2014

Report

Report Number
2649622-2014-09656
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 21, 2014
Report Date
June 10, 2014
Manufacturer
MPRI
Product Code
DXY
PMA / PMN Number
P890003
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE. ELECTRICAL RESISTANCE AND CROSS CONTINUITY TEST RESULTS WERE WITHIN SPECIFICATION.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING. THE DEVICE AND LEAD WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470995 CAPSURE VDD 2 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY MPRI 5038-58

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R ADVDD01, IPG