SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2014-09652
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. DRIED TISSUE/BODY FLUID IN THE SLEEVEHEAD PREVENTS THE HELIX FROM EXTENDING AND RETRACTING. THIS IS NOT A LEAD FAILURE.
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAD HIGH THRESHOLDS DUE TO POSSIBLE DISLODGEMENT. THE RA LEAD WAS REMOVED AND REPLACED. DURING THE SAME IMPLANT PROCEDURE, A NEW RIGHT VENTRICULAR (RV) LEAD WAS ATTEMPTED AND NOT IMPLANTED AS THE HELIX WOULD NOT EXTEND. ANOTHER RV LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470992 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M62 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00055 YR |