FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3992807 · Received August 8, 2014

Report

Report Number
2649622-2014-09652
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. DRIED TISSUE/BODY FLUID IN THE SLEEVEHEAD PREVENTS THE HELIX FROM EXTENDING AND RETRACTING. THIS IS NOT A LEAD FAILURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAD HIGH THRESHOLDS DUE TO POSSIBLE DISLODGEMENT. THE RA LEAD WAS REMOVED AND REPLACED. DURING THE SAME IMPLANT PROCEDURE, A NEW RIGHT VENTRICULAR (RV) LEAD WAS ATTEMPTED AND NOT IMPLANTED AS THE HELIX WOULD NOT EXTEND. ANOTHER RV LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470992 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1 00055 YR