FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3992793 · Received August 8, 2014

Report

Report Number
2649622-2014-09681
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 19, 2014
Report Date
June 19, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND, HOWEVER THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT, THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS, AND VISUAL SUMMARY ANALYSIS OF THE LEAD I NDICATED DAMAGE AT IMPLANT; FULL LEAD RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THERE WERE DIFFICULT, MULTIPLE ATTEMPTS TO POSITION THE LEAD, INCLUDING REMOVING IT TO RESTICK. THE PHYSICIAN REQUESTED A NEW LEAD AFTER FAILED ATTEMPTS OF THE HELIX TO AFFIX. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470809 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1 00037 YR