FDA Adverse Event Injury Summary report: N

ENTRUST

MDR report key: 3992791 · Received August 8, 2014

Report

Report Number
3004209178-2014-14765
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 12, 2014
Report Date
June 12, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Removal / Correction Number
Z-1249-2012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, RETURNED PRODUCT TESTING FOUND THE DEVICE DID PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE RETURNED DEVICE INDICATED SHORTING/LOW IMPEDANCE IN THE BATTERY.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD REACHED ERI (ELECTIVE REPLACEMENT INDICATOR) AND AT THE TIME OF GENERATOR REPLACEMENT THE DEVICE WAS AT A DEPLETED CONDITION WITH NO TELEMETRY. IT WAS NOTED THAT THE DEVICE WAS AT NORMAL ERI BUT DEFERRED REPLACEMENT FOR AN EXTENDED TIME. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469053 ENTRUST DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154VRC

Patients

Seq Age Sex Outcome Treatment
1 00067 YR Hospitalization| R 694765 LEAD