FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3992767 · Received August 8, 2014

Report

Report Number
2649622-2014-09692
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 22, 2014
Report Date
May 22, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE PROXIMAL CONDUCTOR WAS DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS AND VISUAL ANALYSIS NOTED THE LEAD WAS DAMAGED AN IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS ATTEMPTED TO BE IMPLANTED BUT NOT USED DUE TO HIGH IMPEDANCES MEASURED THROUGH THE ANALYZER AT SEVERAL DIFFERENT LOCATION PLACEMENTS. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468997 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00076 YR