FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX MRI SURESCAN
MDR report key: 3992767
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-09692
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 22, 2014
- Report Date
- May 22, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE PROXIMAL CONDUCTOR WAS DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS AND VISUAL ANALYSIS NOTED THE LEAD WAS DAMAGED AN IMPLANT. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD WAS ATTEMPTED TO BE IMPLANTED BUT NOT USED DUE TO HIGH IMPEDANCES MEASURED THROUGH THE ANALYZER AT SEVERAL DIFFERENT LOCATION PLACEMENTS. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 468997 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00076 YR |