FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3992763 · Received August 8, 2014

Report

Report Number
2649622-2014-09671
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD, THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH, BLOOD WAS OBSERVED ON THE SLEEVEHEAD OF THE LEAD AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WITHIN A DAY AFTER THE RIGHT VENTRICULAR (RV) LEAD IMPLANT, THE SLACK HAD BEEN PULLED OUT AND THE RV LEAD WAS PULLED BACK. IT WAS ALSO REPORTED THAT THERE WAS INTERMITTENT RV CAPTURE AND UNDERSENSING OF THE PATIENT'S NATIVE RHYTHM. THE RV LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470713 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00079 YR Hospitalization| R