FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3992758 · Received August 8, 2014

Report

Report Number
2649622-2014-09695
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: DTBA1D4 ICD, IMPLANTED: (B)(6) 2014; 429678 LEAD, IMPLANTED: (B)(6) 2014. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING LOWER HEART RATES IN THE PRESENCE OF T-WAVE OVERSENSING (TWOS) ON THE RIGHT VENTRICULAR (RV) LEAD. LEAD SENSITIVITY AND PACING CONFIGURATION WERE ADJUSTED WITHOUT RESOLVE. THE LEAD WAS THEN REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469756 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M55

Patients

Seq Age Sex Outcome Treatment
1 00064 YR Hospitalization| R 5086MRI45 LEAD