FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3992729 · Received August 8, 2014

Report

Report Number
2649622-2014-09711
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
March 11, 2014
Report Date
June 17, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION.  BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. CONCOMITANT MEDICAL PRODUCTS: 4076 LEAD, IMPLANTED: (B)(6) 2011. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) TRIGGERED, AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING AND NOISE. THE LEAD WILL CONTINUE TO BE MONITORED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470450 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694758

Patients

Seq Age Sex Outcome Treatment
1 00062 YR D284DRG ICD