FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3992719 · Received August 8, 2014

Report

Report Number
2649622-2014-09704
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 1, 2014
Report Date
May 1, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076-58 LEAD, IMPLANTED: (B)(6) 2013; 5554-53 LEAD, IMPLANTED: (B)(6) 2008. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SIGNIFICANT TRICUSPID REGURGITATION THAT WAS ATTRIBUTED TO THE PACEMAKER LEADS. TO IMPROVE THE PATIENT'S OVERALL FUNCTION, THE PHYSICIAN OPTED TO REMOVE THE PACEMAKER SYSTEM. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469520 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00062 YR Hospitalization| R ADDRL1 IPG