FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 3992705 · Received August 8, 2014

Report

Report Number
2649622-2014-09718
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0070-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4)

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NO CAPTURE AND CAPTURE THRESHOLD THAT HAD BEEN CONSISTANTLY RISING OVER TIME. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC WHEN THE LEAD INTEGRITY ALERT (LIA) WAS TRIGGERED DUE TO ELEVATED SHORT INTERVAL COUNTS (SIC), HIGH NON-SUSTAINED TACHYCARDIA (NST) EPISODES AND NOISE DUE TO A FRACTURE. THE RV LEAD WAS PROGRAMMED OFF AND WAS LATER CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470353 SPRINT FIDELIS DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694965

Patients

Seq Age Sex Outcome Treatment
1 00056 YR Hospitalization| R (B)(4)